Development and Validation by RP-HPLC Method for Estimation of Zidovudine in Bulk and Its Pharmaceutical Dosage Form

 

M.A.Haneef1, B.Rajkamal1, G. Tharun goud2, V. Mohan Goud2

1Ganga Pharmacy College, Nizamabad, A.P. India.

2University College of Pharmaceutical Sciences, Palamur University, Mahabubnagar.

*Corresponding Author E-mail: mohanvanga@yahoo.com

 

 

ABSTRACT:

A rapid and precise reverse phase high performance liquid chromatographic method has been developed for the validated of zidovudine, in its pure form as well as in tablet dosage form. Chromatography was carried out on a symmetry C18 (4.6 x 150mm, 5µm) column using a mixture of, methanol and phosphate buffer (65:35 v/v) as the mobile phase at a flow rate of 1.0ml/min, the detection was carried out at 264nm. The retention time of the drug was 2.55±0.02min.The method produce linear responses in the concentration range of 20-80mg/ml of zidovudine. The method precision for the determination of assay was below 2.0%RSD. The method is useful in the quality control of bulk and pharmaceutical formulations.

 

KEYWORDS: Zidovudine, RP-HPLC, validation.

 

 


INTRODUCTION:

Zidovudine[1,2] (INN) or azidothymidine (AZT) (also called ZDV) is a nucleoside analog reverse-transcriptase inhibitor (NRTI), a type of antiretroviral drug used for the treatment of HIV/AIDS infectiousness. It is chemically 1-[(2R,4S,5S)-4-azido-5-(hydroxymethyl)oxolan-2-yl]-5-methylpyrimidine-2,4-dione.The author have developed a liquid chromatographic and validated, sensitive and reproducible method for the determination of zidovudine in bulk and pharmaceutical dosage forms.

 

Chemical structure of zidovudine:

 

 

Received on 16.03.2013       Modified on 26.03.2013

Accepted on 29.03.2013      © AJRC All right reserved

Asian J. Research Chem. 6(4):  April 2013; Page   341-344

MATERIALS AND METHODS:

Chromatographic conditions

A perominence isocratic HPLC system (waters high performance liquid chromatography with auto sampler and UV detector) column symmetry C18 (4.6X150mm, 5µm). A 20µL Rheodyne injection syringe was used for sample injection. HPLC grade, methanol and phosphate buffer were used for the preparing the mobile phase. A freshly prepared Methanol: 0.05M potassium di-hydrogen phosphate buffer (pH-2.8) (65:35 v/v) was used as the mobile phase. The solvents was filtered through a 0.45µ membrane filter and sonicated before use. The flow rate of the mobile phase was maintained at 1mL/min., column temperature was maintained at room temperature and the detection of the drug was carried out at 215nm.

 

Preparation of phosphate buffer

Weigh 7.0 grams of potassium di hydrogen phosphate into a 1000ml beaker, dissolve and diluted to 1000ml with HPLC water. Adjusted the pH to 2.8 with orthophosphoric acid.

 

Preparation of mobile phase

Mix a mixture of above buffer 350mL (35%) and 650mL of Methanol HPLC (65%) and degas in ultrasonic water bath for 5 minutes. Filter through 0.45µ filter under vacuum filtration.

 

Diluent preparation:

Mobile phase as diluents.

 

Standard solution preparation:

Accurately weigh and transfer 10mg of zidovudine working standard in to 10mL volumetric flask add about 10ml of diluent and sonicated to dissolve it completely and the make volume up to the mark with the same solvent. (Stock solutions) further pipette 0.4ml of the above stock solution into a 10ml volumetric flask and dilute up to the mark with diluent. Mix well and filter through 0.45 µm filter.

 

Sample solution preparation:

Weigh 5 zidovudine tablets and calculate the average weight. Accurately weigh and transfer the sample equivalent to 10mg of zidovudine into a 10ml volumetric flask. Add about 10ml of diluent and sonicated to dissolve it completely and make volume up to the mark with diluent. Mix well and filter through 0.45µm filter. Further pipette 0.4ml of the above stock solution into a 10ml volumetric flask and dilute up to the mark with diluent. Mix well and through 0.45µm filter.

 

METHOD VALIDATION:

Linearity:

The linearity of the method was demonstrated over the concentration range of 20-80 mcg/ml of the target concentration. Aliquots of 20, 30, 40,50,60,70 and 80 µg/ml were prepared from above stock solution. Different concentrations of the pure drug were injected into the chromatographic system. Calibration curve of zidovudine was constructed by plotting peak area versus applied concentration of zidovudine [3-10]. A typical chromatogram is shown in Fig 1. The obtained results shown an excellent correlation between peak area and concentration of pure drug within the concentration range and it has shown in fig: 2. The correlation coefficient for the average area at each level versus concentration of analyte was calculated and is presented in Table: 1 and their calibration parameters were shown in Table: 2.

Precision method

The precision of the method was demonstrated by inter-day and intra-day variation studies. In the intra-day studies, six repeated injections of standard solution was made and the response factor of drug peak and% RSD were calculated and present in Table 3. The chromatogram was shown in Fig 3. In the inter-day variation studies, six repeated injections of standard solution were made for six consecutive days and response factor of drug peak and %RSD were calculated shown in Table3. From the data obtained, the developed method was found to be precise.

 

Accuracy

A study of recovery of zidovudine from spiked placebo was conducted at three different spile levels i.e.50%, 100% and 150% samples were prepared with zidovudine raw material equivalent to about the target initial concentration of zidovudine. Sample solutions were prepared in triplicate for each spike level and assayed as per proposed method. The % recovery was given in Table 4. The mean recoveries of zidovudine from spiked were found to be in the range of 98.72-100.65%.

LOD and LOQ:

The LOD and LOQ were separately determined based on the standard deviation of response of the calibration curve. The residual standard deviation of the regression lines and slope of the calibration curves were used to calculate the LOD and LOQ (Table no.2)

 

System suitability

System suitability tests are an integral part of method development and are used to ensure adequate performance of the chromatographic system. Retention time (Rt), number of theoretical plates (N) and tailing factor (T) were evaluated for six replicate injections of the drug at a concentration of 40µg/ml. The results given in Table5 were within acceptable limits.

 

 


Table 1. Linearity results for zidovudine

Conc. (µg / ml)

20

30

40

50

60

70

80

Avg. area

585.6588

896.5403

1198.85

1521.78

1825.699

2135.63

2739.793

Correlation

0.999

 

Table:2 Characteristic parameters of zidovudine for the proposed RP-HPLC method

Parameters

RP-HPLC

Calibration range (mcg/ml)

20-80

Detection wavelength

264nm

Mobile phase (Methanol: Buffer)

65:35

Retention time

2.566±0.02

Regression equation(Y*)

Y=30.63x-15.32

Slope (b)

30.63

Intercept (a)

15.32

Correlation coefficient (r2)

0.999

Intraday precision (%RSD*)

Interday precision (% RSD*)

0.16

0.27

Limit of detection ( mcg/ ml)

0.27

Limit of quantitaion(mcg/ml)

0.83

 

Table 3. Precision results for zidovudine

Sr no

Concentration (µg /ml)

Intraday precision (area)

Interday precision (area)

1

40

1239.961

1240.952

2

40

1238.251

1241.183

3

40

1241.653

1242.864

4

40

1242.721

1245.744

5

40

1238.129

1249.813

6

40

1242.218

1245.813

Mean

 

1240.489

1244.395

Std Dev

 

2.009139

3.397069

% RSD

 

0.16

0.27

 

Table 4. Accuracy results for zidovudine

Sample No

Spike level

Amount (µg / ml) added

Amount (µg / ml) found

% Recovery

Mean % Recovery

1

50%

19.8

20

99.00

99.53

50%

19.89

20

99.47

50%

20.00

20

100.12

2

100%

40.76

40

101.9

100.65

100%

39.62

40

99.05

100%

40.41

40

101.02

3

150%

59.09

60

98.48

98.72

150%

59.40

60

98.98

150%

59.23

60

98.72

 

 


Table:5 system suitability studies of zidovudine by RP-HPLC method

Property

Values

Required limits

Retention time (Rt)

2.566±0.02

RSD ≤ 1%

Theoretical plates (N)

2156.8

N > 2000

Tailing factor

1.3

T≤ 2

 

Fig: 1. Chromatogram of zidovudine at 264nm

 

Fig:2. Calibration curve of zidovudine

 

Fig:3. Chromatogram of precision

 

RESULTS AND DISCUSSION:

In HPLC method, HPLC conditions were optimized to obtain, an adequate separation of eluted compounds. The objective of this study was to develop a rapid and sensitive RP-HPLC method for the analysis of zidovudine in bulk dug and pharmaceutical dosage form by using the most commonly employed RP C-18 column with UV-detection.

The run time was set at 6min and the retention time for zidovudine was 2.56±0.2 min. Each sample was injected 5 times and the retention times were same. When the concentrations of zidovudine and its respective peak areas were subjected to regression analysis by least squares method, a good linear relationship (r2=0.999) was observed between the concentration of zidovudine and the respective peak areas in the range 20-80 mcg / ml. The regression equation was used to estimate the amount of zidovudine, either in tablet formulations or in validation study (precision and accuracy). For the proposed RP-HPLC method, characteristic parameters were shown in Table: 2.

To analyse tablet formulations, RP-HPLC method has been developed. Zidovudine tablets were analyzed as per the procedure described above. The low % RSD values (≤ 2) indicated that the method was precise and accurate. The mean recoveries found in the range of 98.72 – 100.65%. No interfering peaks were found in the chromatogram indicating that excipients used in the tablet formulation did not interfere with the estimation of the drug by the proposed RP-HPLC method.

 

CONCLUSION:

The proposed RP-HPLC method was also validated for intra and inter-day variation. When the solution containing 40 mcg/ml of zidovudine was repeatedly injected on the same day, the % RSD in the peak area for six replicate injections was found to be 0.16%. Also the inter day variation (6 days and six injections) was found to be 0.27%. The results are presented in Table:3. The % RSD values were within 2 and the method was found to be precise. It can concluded the proposed HPLC method is rapid, sensitive, precise and accurate for the determination of zidovudine and can be reliably adopted for routine quality control analysis of zidovudine in Bulk and its pharmaceutical formulations.

 

ACKNOWLWDGEMENT:

I would like to thank Aurobindo Pharma, Hyderabad for providing reference sample of zidovudine and thank my Principle Dr. B. Rajkamal, Ganga Pharmacy College, Nizamabad for providing facilities for my work and specially I would like to thank G. Tharun  Goud, Head, Department of Pharmaceutical Sciences, Palamur University, Mahabubnagar for guiding me in my project work.

 

REFERENCES:

1.       Anonymous 1.URL: http://en.wikipedia.org/wiki/Zidovudine

2.       Anonymous 2.Indian pharmacopoeia-2007, volume-3.

3.       Meyer VR. Practical high performance liquid chromatography. 2nd ed. London: John Wiley and Sons; 1993.

4.       Kazakevich Y, Lobrutto R.HPLC for pharmaceutical scientists. London: John Wiley and Sons; 2007.

5.       ICH: Q2B, Analytical Validation- Methodology, 1996.

6.       David M, Bliesner. Validating Chromatographic Methods - A Practical Guide - New Jersey: John Wiley and Sons Inc.; 2006.p. 1, 9-12.

7.       Paul C Sadek. The HPLC Solvent Guide. 2nd Ed. New York: Wiley-Interscience John Wiley and Sons Inc.; 2007. p. 67-74

8.       Ermer J, Miller JH Mc; B. Method Validation in Pharmaceutical Analysis-A Guide to Best Practice. Weinheim: Wiley-Vch Verlag GmbH and Co. KgaA; 2005. p. 21, 52, 63, 80, 99, 101, 120, 170.

9.      Hokanson GC. A life cycle approach to the validation of analytical methods during pharmaceutical product development.

10.     Riley CM, Thomas WR. Development and validation of analytical methods. Vol.III. United states of America: Elsevier Sciences Inc; 1996.

 

 

 

Received on 11.03.2013                             Modified on 25.03.2013

Accepted on 29.03.2013                           © AJRC All right reserved

Asian J. Research Chem. 6(4):  April 2013; Page   334-340